Reveal® Q+ for DON EndPoint

25 tests per kit

SKU No.  700002536  |  Catalog No.  8375

  • For use with the Raptor® Solo Platform
  • Uses the standard track mode
  • Easy to use for increased workflow output
Qty:
Minimum quantity required: 1
This item must be ordered in multiples of one case. Quantity per case: 1
Reveal® Q+ for DON EndPoint is an easy-to-use quantitative lateral flow test based on a competitive immunoassay format. The assay is read on the Raptor® Solo Platform using the standard mode track. The EndPoint format is designed for use when increased workflow output is needed.
Official_Specifications-Lateral-Flow-Device_Illustration.png
Specifications
BrandReveal®
AnalyteDeoxynivalenol (DON)
PlatformLateral Flow
Result TypeQuantitative
ApprovalsFGIS: 2021-154
Quantity per Package25 tests
Package Dimensions8.50 in x 2.50 in x 2.50 in
Package Weight0.47 lb
  1. 25 Reveal® Q+ for DON EndPoint test strips
  2. 25 red sample dilution cups
  3. 25 clear sample dilution cups
  4. 1 bottle of sample diluent
  1. Agri-grind grinder or equivalent (Neogen items 9401, 9453)
  2. Scale capable of weighing 5–50 g ± 0.1 g (Neogen item 9427)
  3. Timer (Neogen items 9426, 9452)
  4. Reveal sample cup rack (Neogen item 9475)
  5. Pipettor, 100 µL (Neogen items 9272, 9860)
  6. Pipette tips 1-200 µL (Neogen items 9407,9410)
  7. Pipettor 500 µL fixed (Neogen item 9336)
  8. Pipette tips 100-1,0000 µL (Neogen items 9464, 9487)
  9. Mini centrifuge (Neogen item 9330)
  10. Micro centrifuge tubes (Neogen item 9172)
  11. Filter syringes (Neogen item 9420)
  12. Distilled or deionized water
  13. Sample collection cups (Neogen item 9428)
  14. Sample collection tubes (Neogen items 9421, 9428)
  15. Graduated cylinder 50 mL (Neogen item 9447)
  16. Raptor Solo reader (Neogen item 9696)

Technology

In the test, the extract is wicked through the reagent zone, which contains antibodies specific for DON conjugated to colloidal gold particles (gold complex). If DON is present, it will be captured by the gold complex. The gold complex, along with any free gold-complex, is then wicked onto a membrane, which contains a zone of DON conjugated to a protein carrier. This zone captures any unbound gold complex, allowing the particles to concentrate and form a visible line. As the level of DON in a sample increases, free aflatoxin will bind with the gold complex allowing the less gold complex to be captured in the test zones. Therefore, as the concentration of DON in the sample increases, the test line density decreases.

Algorithms programmed into the readers convert the line intensities into a quantitative result displayed in parts per billion (ppb). The membrane also contains a control zone where an immune complex present in the reagent zone is captured by an antibody, forming a visible line. The control line will always form regardless of the presence of DON, ensuring the strip is functioning properly.

Official_BetaStar_Technology-Photo.png

Training

Our customers’ success is our shared success. Our experts are ready to train you and your team on our solutions, so you can rest easy knowing procedures are performed properly and yield accurate results. In addition, we provide certificates upon completion of training to provide you with the documentation needed for audit traceability.

Support

Validation

Our experts are here to help. If you need information beyond what is provided in Validation Reports under the Documents tab, please contact us and we will be happy to assist you. Learn more