Testosterone ELISA Kit

96 well

SKU No.  402510  |  Catalog No. 

  • Highly sensitive
  • Results in less than two hours
  • Accurate and easy to use
Qty:
Minimum quantity required: 1
This item must be ordered in multiples of one case. Quantity per case: 1

Testosterone ELISA is used for the quantitative analysis of testosterone levels in biological fluid.

Testosterone is the principle androgen. It is synthesized in the testis, the ovary, and the adrenal cortex. It is responsible for the development and maintenance of the male secondary sex characteristics. It also exerts important protein anabolic and growth-promoting effects. The plasma testosterone levels are useful in investigating hypogonadism and hormone replacement therapy in men. It is also useful as a marker in hyperandrogenism in women.

Specifications
Brand Neogen®
Analyte Testosterone, Dihydrotestosterone
Application Urine, Oral Fluid, Tissue Culture Supernatant, Plasma, Serum
Assay Range 0.002 - 0.2 ng/mL
Intended Use For the determination of Testosterone in urine, oral fluid, tissue culture supernatant, plasma, or serum samples. Contact a Neogen representative for information on other matrices.
Species Non-Species Specific
Assay Sample Size 50 µL
Platform ELISA
Result Type Quantitative
Microplate Size 96 well microplate
Total Assay Incubation Time 90 minutes
Drug Panel Hormones & Steroids
Storage Conditions 2-8°C
Wavelength 650 nm or 450 nm with acid stop
Package Weight 0.70 lb
Analytes Cross-Reactivity
Testosterone 100.0%
Dihydrotestosterone 100.0%
Testosterone Glucuronide 16.12%
Androstenedione 0.86%
Bolandiol 0.86%
Testosterone Enanthate 0.13%
Estriol 0.10%
Testosterone Benzoate 0.10%
Estradiol 0.05%
Dehydroepiandrosterone 0.04%
Testosterone Propionate 0.04%
Deoxycorticosterone 0.03%
Testosterone 17β-Cypionate 0.02%
Aldosterone < 0.01%
Corticosterone < 0.01%
Cortisol < 0.01%
Cortisone < 0.01%
Estrone < 0.01%
17-Hydroxyprogesterone < 0.01%
Pregnenolone < 0.01%
Progesterone < 0.01%
  1. EIA Buffer
  2. Wash Buffer (10X)
  3. K-Blue Substrate
  4. Extraction Buffer (5X)
  5. Testosterone Enzyme Conjugate
  6. Testosterone Standard
  7. Testosterone Antibody Coated Microplate
  1. Deionized water
  2. Precision pipettes that range from 10 μL-1000 μL and disposable tips. NOTE: If all or several strips are to be used at one time, it is suggested that a multichannel pipette be used.
  3. Clean test tubes used to dilute the standards and conjugate.
  4. Graduated cylinders to dilute and mix wash buffer and extraction buffer.
  5. Microplate reader with 650 nm filter.
  6. Plate cover or plastic film to cover plate during incubation.

Additional Information

Calculation of Results

Neogen has partnered with MyAssays.com to provide an online tool for data calculations.

Follow the simple step-by-step guide to generate your data.

Calculate Results

MyAssays.com is a service of MyAssays Ltd. For questions or concerns about the service, please contact MyAssays Ltd.

Technology

The sample or standard solution is first added to the microplate. Next, the diluted enzyme conjugate is added, and the mixture is shaken and incubated at room temperature for one hour. During the incubation, competition for binding sites is taking place. The plate is then washed removing all the unbound material. The bound enzyme conjugate is detected by the addition of substrate which generates an optimal color after 30 minutes. Quantitative test results may be obtained by measuring and comparing the absorbance reading of the wells of the samples against the standards with a microplate reader at 650 nm. The extent of color development is inversely proportional to the amount of Testosterone in the sample or standard. For example, the absence of Testosterone in the sample will result in a bright blue color, whereas the presence of Testosterone will result in decreased or no color development.
AssayPrincipleIllustration_LifeScience.png

Training

Our customers’ success is our shared success. Our experts are ready to train you and your team on our solutions, so you can rest easy knowing procedures are performed properly and yield accurate results. In addition, we provide certificates upon completion of training.

Support

The following are common examples of unexpected results when running Neogen's ELISA kits with possible explanations.

  1. Extreme deep blue color development with samples and standards.
    1. The plate was not properly washed (3 x 300 µL) with the diluted wash buffer. If using an automated washer, ensure the machine is working correctly and increase wash cycle to 5 x 300 µL.
    2. The enzyme conjugate concentrate was incorrectly diluted.

  2. Extremely low color development with samples and standards.
    1. The wash buffer was not diluted correctly before use.
    2. A poor quality of water was used to dilute the wash buffer. Deionized water should be used for dilutions.
    3. The enzyme conjugate concentrate was incorrectly diluted.
    4. The kit prematurely deteriorated, possibly from adverse shipping and storage conditions. Investigate the condition of the kit when it was received and how it was stored prior to use. Proper storage conditions are listed in the package insert.
    5. The kit has expired. Check the expiration dates on the test kit and reagents. No component of the kit should be used past the expiration date. Do not mix any reagents or components of one kit with the reagents or components of another kit.
    6. Contamination. Always use aseptic techniques when opening and removing reagents from vials and bottles. Keep the plate covered except when adding reagents, washing or reading. Always use different pipette tips for each reagent. When pipetting, do not allow the pipette tip to touch any of the reagents already in the well.

  3. No color development with samples and standards.
    1. Improper dilution of enzyme conjugate concentrate.
    2. The kit has expired. Check the expiration dates on the test kits and reagents. No components of the kit should be used past the expiration date. Do not mix any reagents or components of one kit with the reagents or components of another kit.
    3. Incorrect addition of kit reagents. Review assay procedure.

  4. Little to no displacement with the standard curve.
    1. Incorrect dilution of standards. Refer to the dilution scheme in the package insert.
    2. Contamination.
    3. The plate was not properly washed (3 x 300 µL) with the diluted wash buffer. If using an automated washer, ensure the machine is working correctly and increase wash cycle to 5 x 300 µL.
    4. Standard has deteriorated prematurely. Contact your Neogen representative and provide them with the kit name, lot number, expiration date and your OD readings for further investigation.

  5. The standard curve performed correctly, but the known negative samples gave low color development.
    1. Samples need to be diluted or extracted to eliminate interference. Refer to the extraction procedure in the kit insert.

  6. The standard curve performed correctly, but the extracted samples gave low color development.
    1. The concentration of the analyte in the extracted sample is too high. The extracted sample will need to be diluted before running it in the assay so the sample OD reading will fit in the standard curve. When a dilution is used, the concentration determined from the standard curve must be multiplied by the dilution factor.
    2. Inadequate extraction of samples resulting in the presence of a solvent. Refer to the recommended extraction procedure in the package insert.

  7. Variability with duplicates.
    1. Inconsistent and/or inadequate pipetting technique when adding reagents. Improve pipetting technique.
    2. Inconsistent washing. The wells should be washed 3 x 300 µL with the diluted wash buffer. If using an automated washer, ensure the machine is working correctly and increase wash cycle to 5 x 300 µL.
    3. Inadequate aspiration during washing. The wells should be dumped and tapped between each washing. Tap out excess liquid in wells before adding substrate. Add substrate immediately to the wells once the excess wash buffer has been removed. If using an automated washer, ensure the machine is aspirating correctly.
    4. Interruption during assay set-up. Have all reagents prepared before assay set-up commences. Reagent addition should be performed in a timely and accurate manner.